Aethalometer
Model AE33
Chamber
6
Filter tape
8 and 7
Setup file
9
Tests &
12, 11, 10
procedures
External
13
device
Auto clean
14
air test
status
CF card
15
failure
Database
16
status
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
0
0
Chamber OK
1
64
Chamber error
00
0
Filter tape OK
01
128
Tape warning (less than 30 spots left)
10
256
Tape last warning (less than 5 spots left)
11
384
Tape error (tape not moving, end of tape)
0
0
Setup file OK
1
512
Setup warning
000
0
No test
001
1024
Stability test
010
2048
Clean air test
011
3072
Change tape procedure
100
4096
Optical test
110
6144
Leakage test
0
0
Connection OK
1
8192
Connection Error
0
0
Status OK or test not run yet
1
16384
Result of clean air test is not acceptable
Service of the instrument is recomended
1
32768
Problem while saving or retrieving files
to/from CF card
0
0
Database status OK
1
65535
Database bigger than 2*106 data lines
July 2017