10. SERIAL DATA PORT CONNECTIONS

All commands finish with a Carriage Return. Note that all commands must be sent as
UPPERCASE characters.
$AE33:Dnnn
$AE33:Wnnn
$AE33:TyyyyMMddHHmmss
$AE33:A
$AE33:SG
$AE33:SSn m
$AE33:Lnnn
$AE33:Fn m
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
nnn = number of lines of
data requested
nnn = number of lines of
data requested
yyyy
year
MM
month
dd
day
HH
hour (24)
mm
min
ss
second
n = parameter name
m = parameter value
nnn = number of log
entries requested
n = folder name (Data,
NDtest, CleanAirTest,
StabilityTest,
July 2017
Request nnn last data lines
Request nnn last
measurements.
Same as $AE33:Dnnn, except
that BC
(ng/m3) is reported
BB
instead of BB (%)
Synchronize time with an
external device (e.g.
datalogger)
Report tape advances
remaining
Get setup file
Set setup parameter
Example: $AE33:SSTimebase 1
Note: instrument must be
stopped
Request nnn last log entries
Get file(s) – starts streaming
number of files