Model AE33

10.2 Connecting to External Datalogger or PC

Connect with a crossover cable (or standard serial cable with null modem adapter) from the
PC / Datalogger to the serial port of the instrument on the rear panel.
Establish communications using the following values:
BaudRate = 115200
DataBits = 8
StopBits = 1 (IO.Ports.StopBits.One)
Parity = None (IO.Ports.Parity.None)
Handshake = None (IO.Ports.Handshake.None)
DTR_Enable = True
RTS_Enable = True
Note: the BH and Qair protocols use these connection parameters above.