Aethalometer
Model AE33
o Automatic time adjustment: None, NTP or GPS. NTP uses "Network Time
Protocol" to synchronize time between computers. Instrument must be
connected via LAN to accessible NTP server. For GPS time adjustment a GPS
must be connected to the instrument as an external device (GPS_GSTAR_IV (PN
AE33-GPS)). Time is adjusted during the tape advance procedure.
 ADVANCED screen shows all the parameters used internally:
Status
Controller
status
Detector status
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
The same value as Status reported on the Home screen (see
section 13.2 Instrument status).
0 – ready for operation; 100 – reset detected, not read for
operation; 255 – busy
0 – Detector initialization ERROR; 10 – Detector Measure; 20
– Detector Stop; 30 – LED calibration; 40 – Fast LED
calibration; status numbers 55 – 58 indicate detector is
July 2017
46/146