Aethalometer
Model AE33

8.2 Instrument Status

The Instrument Status is shown by a status flag constructed as a 16-bit value composed of
codes for various instrument functions. These include the Detector Status (bits 0,1); Flow (bits
2,3); Optical Source (bits 4,5); Chamber Position (bit 6); Tape Sensors (bits 7,8); Ball Valve (bit
9); Test Procedures (bits 10, 11 and 12); and other functions as shown in the table below.
The overall Status Code Number is the sum of all the decimal values of all of the possible
status flag values.
Status
bit
relates to:
position
Operation
1 and 0
Flow
3 and 2
Optical
5 and 4
Source
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
status flag
description
binary decimal
00
0
Measurement
01
1
Tape advance (tape advance, fast
calibration, warm-up)
10
2
First measurement – obtaining ATN0
11
3
Stopped
00
0
Flow OK
01
4
Flow low/high by more than 0.5 LPM or
F1 < 0 or F2/F1 outside 0.2 – 0.75 range
10
8
Check flow status history
11
12
Flow low/high & check flow status
history
00
0
LEDs OK
01
16
Calibrating LED
10
32
Calibration error (at least one channel
OK)
11
48
LED error (all channels calibration error,
COM error)
July 2017
56/146