Aethalometer
Model AE33
2. Go to the "Operation / General" screen. If the measurements are running, press 'Stop'.
3. Press the "Verify flow" button to start the test.
4. A new screen will open. Choose "Manual". The "Auto" option can be only used if you are
using the TSI 4110 mass flow meter connected by data cable to the AE33.
5. The first part of the routine is to measure the flow through the flow calibration pad.
6. Wait for the chamber to lift: when prompted, remove the filter tape and install the rubber
flow calibration pad with the notch facing towards you.
7. When installed, press "OK".
8. Connect your external air mass flow meter to the inlet port.