Aethalometer
Model AE33
loading'; as well as a real-time output of the 'loading compensation parameter' k which provides
insights into the aerosol nature and composition. This process is performed in real time for all
wavelengths: examination of the 'k' values as a function of wavelength provides further
information about the aerosol composition. An example of this is shown for extreme
concentrations of black carbon (Figure 8).
Figure 8. The time-series of AE33 raw and compensated BC concentrations with 1 second timebase – note
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
80000
60000
40000
20000
0
2000
the extreme concentrations and loading effects.
IR, 1 second
raw BC1
raw BC2
compensated BC
3000
time (sec)
July 2017
4000
135/146