16. The test is satisfactory if the difference of flow readings is less than +/-10%.
If the difference is larger, re-calibration of the flow sensors is needed (see section 9.3).
17. Confirm by pressing the "OK" button.
18. When prompted, remove the rubber flow calibration pad and re-install the filter tape.
Press "OK" when done.
19. The test is now complete, and normal operation can resume.