sample line tubing with subsequent damage to the instrument and serious effects on the validity
of the measurement data.
The equipment package includes a selection of inlet fittings. According to your choice of inlet
tubing, thread a suitable fitting into the Sample Inlet port on the rear panel of the Aethalometer
(see Section 4, Fitting set). Use Teflon tape on the thread. If using the flexible static-dissipative
tubing, insert the barbed connector into the black conductive tubing and then push it into the
matching fitting on the rear panel of the instrument.
Figure 4. Fit the inlet connector on the rear panel of the Aethalometer and connect the sampling line
Install the insect screen and water trap assembly in the tubing at a location near the
instrument so that it can be easily seen. Connect the other end of the sample line tubing to the
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
tubing to it.