SAFETY NOTES

Please pay attention to additional safety notes as described in detail in the User Manual, Chapter 2,
„SAFETY NOTES AND LABELS".
THE INSTRUMENT
front view
INSTALLATION
The instrument should preferably be installed indoors. If outdoors, it must be protected from the
weather (sun, rain, wind). It should be operated at ambient temperature from 10 to 40°C. If it is
installed in a small enclosure, this must be ventilated by a fan. Avoid placing the instrument or its
sampling lines in the discharge from air-conditioners.
1) Remove
red
protective caps from the
instrument.
2) Attach the sample inlet line using
fittings and adapters. Avoid sharp bends.
3) Switch the instrument on (make sure
to switch on TWO switches, one on rear panel,
one behind front door).
4) Use the size-selective PM2.5 cyclone inlet to exclude
dust and large particles. Clean the cyclone periodically!
5) Use the Sample Stream Dryer especially in cases of
high indoor ambient humidity.
6) Avoid direct discharge of air-conditioner over
sample lines, or insulate them. (This may cause
water condensation in the tubing).
7) Install insect screen/water trap (provided
with the instrument) close to the rear of the instrument.
The cup must be oriented downwards.
Installation example:
Aethalometer AE33, Sample Stream Dryer, cyclone and MET station
Sample Stream Dryer, Cyclone, MET station - OPTIONAL
Any water intake may damage the instrument.
Damage due to water intake is not covered by the warranty!
Sample inlet
8" color Touch screen
Status indicator LED's
CF card cover
Fan cover
Network connection
USB, COM ports
USB ports
Air exhaust
ON/OFF switch
(rear panel and behind front door)
SAMPLE STREAM DRYER
AE3
3
back view
Avoid direct discharge of air-
conditioner over sample lines.
VACUUM PUMP

Advertisement

loading

Summary of Contents for MAGEE Scientific AE33

  • Page 1: Safety Notes

    VACUUM PUMP The cup must be oriented downwards. Installation example: Aethalometer AE33, Sample Stream Dryer, cyclone and MET station Sample Stream Dryer, Cyclone, MET station - OPTIONAL Any water intake may damage the instrument. Damage due to water intake is not covered by the warranty!
  • Page 2 HOME SCREEN AND USER INTERFACE DESCRIPTION Black Carbon concentration (BC6) UVPM (BC1) or BC biomass burning [%] Measured flow rate (2, 3, 4 or 5 LPM) Time base (1 or 60 seconds) Estimated number of tape advances until the end of tape Status of the instrument (press the icon for more information) START/STOP measurement...
  • Page 3 MAINTENANCE PROCEDURE INTERVAL Procedure (refer to User Manual for details) 1 month 6 months 12 months Check sample inlet flow Inspect sample line tubing Inspect/clean size selective inlet/cyclone/insect screen Verify time and date Inspect/clean optical chamber Validate/calibrate flow, perform leakage test Perform Clean Air/Stability check test Perform ND test Lubricate optical chamber sliders...
  • Page 4 (spec sheets, technical papers, video tutorials) and other resources, we encourage all users to complete the registration form at http://www.mageesci.com/register The Users Group site offers interactive resources at www.mageesci.com/group. CONTACT INFORMATION AEROSOL (globally) T: +386-1-4391-700 support@aerosol.eu Magee Scientific (USA,MEX,CAN) T: +1-510-845-2801 support@mageesci.com...