Troubleshooting; Startup Screen Checks - MAGEE Scientific Aethalometer AE33 User Manual

Hide thumbs Also See for Aethalometer AE33:
Table of Contents

Advertisement

Aethalometer
Model AE33

13. TROUBLESHOOTING

13.1 Startup screen checks

Check
Communication Communication PC to
optical chamber controller
Instrument
Obtain data (serial
data
number) from the optical
chamber controller
Storage
CF card operation
Configuration
Read setting from the
settings
setup file
Valves
Operation of the ball valve
Chamber
Optical chamber
movement test
Pump & Flow
Test if pump is working
Device
Win CE operating system
monitoring
test
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
Description
Error
hardware
problem
hardware
problem
CF card error
Setup file error
ball valve not
moving
locked chamber
hardware error
pump
tube
connections
faulty
application file
July 2017
Solution
check cables
check cables
get new CF card and SW
restore setup file from
one of the older setup
files
check cables
unlock chamber
service needed
service needed
reconnect tubes
get new CF card and SW
115/146