7. Repeat on the other side, so that the cartridge is removed with its two elbow connectors.
8. Remove the elbow connectors from the filter itself, this time pressing the grey ring of the
connector.
9. Take a new filter and attach both elbow connectors to this filter with the same orientation of
the flow indicator arrow.
10. Install the filter by pressing the elbow connectors firmly down onto the sockets.
11. Restart the instrument and continue measurements.
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
July 2017
89/146