Mounting Restrictions - Siemens Pointek CLS200 Digital Operating Instructions Manual

Capacitance switches
Hide thumbs Also See for Pointek CLS200 Digital:
Table of Contents

Advertisement

Angled mounting
Pointek CLS300 cable version:
The cable version is designed for top mounting. The cable suspends vertically so that it
reaches into the process at the desired detection level (high or low detection alarm).

Mounting Restrictions

Note:
• Keep the sensor at least 50 mm (2") away from any nozzle or tank wall.
• If multiple units are used, allow at least 500 mm (20") between them, to prevent
interference.
Side View
(section)
500 mm
(20") min
500 mm
(20") min
Sensors must be 500 mm (20") apart.
7ML19985JJ03
Angle
Multiple Units
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
End View
500 mm
(20") min
Mount diagonally if space is restricted.
Page 21