Filling level status:

Adjust 0 % application range setting
The lower limit of the application range (0%) corresponds to the left limit of the bar graph
in the digital display.
For reliable and accurate detection of the process material, select the example in the
table below that most closely resembles your application, and follow the setup
procedure.
Application
General
applications
Demanding
applications
Interface
detection
1.
Follow the setup procedure that most closely resembles your application.
2.
Open the menu View – Display and select the tab Transducer Block: Discrete Input
(Part 1). Make a note of the Sensor_Value (digits)
3.
Close Display window, and enter the sensor value (recorded in the previous step) to
the Value field:
> Input > Transducer Block: Discrete Input > Range of Application> 0%.
Press Enter: the status field reads Changed.
4.
If no more settings need adjusting, download the data to the device and save the
parameter settings offline (the status fields go blank). Otherwise, continue to adjust
other parameters as required.
7ML19985JJ03
Material
dry solids
low viscosity liquids
hygroscopic / wet solids
high viscosity and high
conductivity liquids
liquid A / liquid B
foam / liquid
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
Setup procedure
Sensor uncovered and a minimum of
100 mm (4") free space all around
Sensor immersed and then uncovered,
but retaining as much build up of
material as possible on the sensor.
Immerse the sensor in the material that
has the lowest relative dielectric
constant.