Use the M button to navigate to Mode 20.
3.
Set up the process to provide the recommended condition for your process from the
table below:
Application
General
Demanding
Interface
detection
Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
4.
Record the count value.
Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
5.
6.
Set the sensor value to approximately 1000 counts above the obtained in step 4.
Press ⇑ to edit the value of each digit, and press ⇓ to move the cursor to the next
digit.
After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate
7.
the procedure and save the new value. The display now shows the new value for
the 100% Application Range setting.
8.
Press M to exit.
CLS300 100% Application Range setting
To adjust the 100% Application Range setting for CLS300:
1.
Install the device in the process.
2.
Use the M button to navigate to Mode 20.
3.
Set up the process to provide the recommended condition for your process from the
table below:
Application
General
Demanding
Interface
detection
Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
4.
Record the count value.
Press ⇑ or ⇓ to enter Edit mode; the sensor value is frozen and ready for editing.
5.
Set the sensor value to 105% of the value obtained in step 4 (1.05 x counts). Press ⇑
6.
to edit the value of each digit, and press ⇓ to move the cursor to the next digit.
Page 42
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
Material
• dry solids
• low viscosity liquids
• hygroscopic / wet solids
• high viscosity and high
conductivity liquids
• liquid A / liquid B
• foam / liquid
Material
• dry solids
• low viscosity liquids
• hygroscopic / wet solids
• high viscosity and high
conductivity liquids
• liquid A / liquid B
• foam / liquid
Setup conditions
sensor fully covered
sensor fully covered
sensor immersed in the material with
the highest dielectric constant
Setup conditions
Material in tank should be at the highest
point it could reach during the normal
process.
Material in tank should be at the highest
point it could reach during the normal
process.
Sensor immersed in the material with
the highest dielectric constant
7ML19985JJ03