Mode 14: Unit (default: °C)
Mode 14 becomes available only after Electronics Temperature is selected in
Note:
Mode 13.
If Electronics Temperature is selected as the display source in Mode 13, you can change
the corresponding physical unit by selecting one of the following options:
Physical Unit
K
°C
°F
°R
To set the physical unit:
Select Mode 14: the primary reading field displays Unit, and the physical unit
appears in the auxiliary reading field.
Press ⇑ or ⇓ to select the desired physical unit.
Press M to save the selection and exit
Mode 15: Node address
Sets the node address of the device on PROFIBUS. The permitted range is from 0 to 126.
We recommend setting the range between 1 and 125.
To set the PROFIBUS node address:
Select Mode 15. The current node address is shown in the primary reading field.
Press ⇑ or ⇓ to open Edit mode.
Enter the node address in the permitted range: press ⇑ to change the value of
each digit, and press ⇓ to move the cursor to the next digit.
Press ⇓ after editing the least significant digit, to terminate the procedure and to
save the value.
Press M to exit.
Note:
• While the device is in cyclic data exchange with a class 1 Master, it is not possible
to change the address. Abort the cyclic communication to change the address.
• Ensure that you set a unique node address to the device. Otherwise, the devices
with the same addresses will cause confusion on the bus.
1.
Applies to PROFIBUS PA applications only. If in standalone application (no network), leave
the node address at the default value.
7ML19985JJ03
1
(default: 126)