CLS300 Digital Switching Point Setup

Use the following procedure to set up the switch point for the CLS300 Digital model.
1.
Set the 0% Application Range Setting (Mode 19):
a. Install the device in the process.
b. Use the M button to navigate to Mode 19.
c. Set up the process to provide the recommended condition for your process from
the table below:
Application
General
Demanding
Interface
detection
d. Press ⇑ or ⇓. The display switches to the current measured sensor value in
counts. Record the count value.
e. Press ⇑ or ⇓ to enter Edit mode; the sensor value is frozen and ready for editing.
f. Set the sensor value to 95% of the value obtained in step 1d (0.95 x counts).
Press ⇑ to edit the value of each digit, and press ⇓ to move the cursor to the next
digit.
g. After editing the least significant digit (or leaving it unchanged), press ⇓ to
terminate the procedure and save the new value. The display now shows the
new value for the 0% Application Range setting.
h. Press M to exit.
2.
Set the 100% Application Range setting (Mode 20):
a. Install the device in the process.
b. Use the M button to navigate to Mode 20.
c. Set up the process to provide the recommended condition for your process from
the table below:
Application
General
Demanding
Interface
detection
Page 48
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
Material
• dry solids
• low viscosity liquids
• hygroscopic / wet solids
• high viscosity and high
conductivity liquids
• liquid A / liquid B
• foam / liquid
Material
• dry solids
• low viscosity liquids
• hygroscopic / wet solids
• high viscosity and high
conductivity liquids
• liquid A / liquid B
• foam / liquid
Setup conditions
Material in tank should be at the lowest
point it could reach during the normal
process.
Material in tank should be at the lowest
point it could reach during the normal
process. The sensor should have the
minimum possible material build up.
Material in tank should be at the lowest
point it could reach during the normal
process.
Setup conditions
Material in tank should be at the highest
point it could reach during the normal
process.
Material in tank should be at the highest
point it could reach during the normal
process.
Sensor immersed in the material with
the highest dielectric constant
7ML19985JJ03