Mode 20: 100% Application Range Setting - Siemens Pointek CLS200 Digital Operating Instructions Manual

Capacitance switches
Hide thumbs Also See for Pointek CLS200 Digital:
Table of Contents

Advertisement

Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
4.
Record the count value.
Press ⇑ or ⇓ to enter Edit Mode; the sensor value is frozen and ready for editing.
5.
6.
Set the sensor value to approximately 300 counts below the value obtained in step 4.
Press ⇑ to edit the value of each digit, and press ⇓ to move the cursor to the next
digit.
After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate
7.
the procedure and save the new value. The display now shows the new value for
the 0% Application Range setting.
8.
Press M to exit.
CLS300 0% Application Range setting
To adjust the 0% Application Range setting for CLS300:
1.
Install the device in the process.
2.
Use the M button to navigate to Mode 19.
3.
Set up the process to provide the recommended condition for your process from the
table below:
Application
General
Demanding
Interface
detection
Press ⇑ or ⇓. The display switches to the current measured sensor value in counts.
4.
Record the count value.
Press ⇑ or ⇓ to enter Edit mode; the sensor value is frozen and ready for editing.
5.
Set the sensor value to 95% of the value obtained in step 4 (0.95 x counts). Press ⇑ to
6.
edit the value of each digit, and press ⇓ to move the cursor to the next digit.
After editing the least significant digit (or leaving it unchanged), press ⇓ to terminate
7.
the procedure and save the new value. The display now shows the new value for
the 0% Application Range setting.
8.
Press M to exit.

Mode 20: 100% Application Range setting

This point corresponds to the right limit of the bar graph in the digital display.
CLS200 100% Application Range setting
To adjust the 100% Application Range setting for CLS200:
1.
Install the device in the process.
7ML19985JJ03
Material
• dry solids
• low viscosity liquids
• hygroscopic / wet solids
• high viscosity and high
conductivity liquids
• liquid A / liquid B
• foam / liquid
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
Setup conditions
Material in tank should be at the lowest
point it could reach during the normal
process.
Material in tank should be at the lowest
point it could reach during the normal
process. The sensor should have the
minimum possible material build up.
Material in tank should be at the lowest
point it could reach during the normal
process.