Mode 10: HW Write Protection (default: OFF, display – –)
When HW Write Protection is enabled, L appears in the Mode display field, in
Note:
normal measuring mode.
Protects parameters from unintentional modification. This prevents parameter changes
both locally via the keypad and remotely via SIMATIC PDM. The parameters are still
readable.
To enable HW Write Protection:
Select Mode 10.
Press ⇑ or ⇓ to activate keylock and to disable operation: L appears in the
numerical value field.
Press M to exit.
To disable HW Write Protection
Select Mode 10.
Hold ⇑ and ⇓ simultaneously for more than 5 seconds: L disappears from the
numerical value field.
Press M to exit.

Mode 13: Display Source (default: 0)

Determines which value will be displayed.
To select the display source from the options shown in the table below:
Select Mode 13.
Press ⇑ or ⇓ to select the value for the desired source.
Press M to save the selection and exit
You can assign the physical unit in Mode 14, if Electronics Temperature is selected as the
display source.
Display Source
From Discrete Input
Function Block: OUT
From Transducer Block:
Primary Value
From Transducer Block:
Sensor Value
From Transducer Block:
electronics temperature:
Electronics temperature
Page 38
Pointek CLS200/300 Digital – OPERATING INSTRUCTIONS
Auxiliary
Value
Reading
0
OUT
1
PRIM
2
SENS
3
TMP E
Displayed
Units
none
Bar graph and
ON or OFF
none
Bar graph and
ON or OFF
none
Bar graph and
counts value
K, °C, °F, or °R
Temperature in
units
7ML19985JJ03