Ph Sensor Systems Connections - ABB AX416 User Manual

Single and dual input analyzers for ph/redox orp
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468

6.6 pH Sensor Systems Connections

Terminal Block B
Sensor
Sensor
B
A
Temperature Compensator (if fitted) Common -
B1
B9
B2
B10
B3
B11
B4
B12
B5
B13
B6
B14
B7
B15
B8
B16
Table 6.1 Non-ABB pH Sensor System Connections
Terminal Block B
7650 and 7660 Without Solution Earth (Ground Rod)
Sensor
Sensor
Color
B
A
B1
B9
Red
B2
B10
Red
B3
B11
White
B4
B12
Not Used
B5
B13
Not Used
B6
B14
Black
B7
B15
Yellow
B8
B16
Clear
Table 6.2 ABB 7650 and 7660 Series pH Sensor Connections
Note.
1. Where pH systems are used with a Solution Earth (ground rod) and those including diagnostics, ensure that
the relevant sensor(s) is set to
2. If the sensor is fitted with a 2-wire Pt100, Pt1000 or Balco 3K temperature compensator, link terminals B9 and B10 (and B1
and B2 if dual input analyzer).
The redox (ORP) version of the 7650 series does not incorporate a temperature sensor. The Temperature sensor must be
disabled in the Configuration page - see section 5.3, page 30.
54
pH/Redox (ORP) Sensor Systems
Without Solution Earth (Ground Rod)
(Standard Input)
see also Note 2 below
Temperature Compensator (if fitted) 3rd Lead
Temperature Compensator (if fitted)
Not Used
Not Used
Reference Electrode
Screen/Shield (if fitted)
Glass/Metal Electrode
Function
Temperature Compensator Common
Temperature Compensator 3rd Lead
Temperature Compensator
Not Used
Not Used
Reference Electrode
Screen/Shield
Glass/Metal Electrode
- see section 5.3, page 30.
Yes
pH/Redox (ORP) Sensor Systems
With Solution Earth (Ground Rod)
(Differential Input - see Note 1 Below)
Temperature Compensator (if fitted) Common -
see also Note 2 below
Temperature Compensator (if fitted) 3rd Lead
Temperature Compensator (if fitted)
Reference Electrode
Not Used
Solution Earth (Ground Rod)
Screen (if fitted)
Glass/Metal Electrode
7660 With Solution Earth (Ground Rod) - see Note 1
below
Color
Red
Temperature Compensator Common
Red
Temperature Compensator 3rd Lead
White
Temperature Compensator
Black
Reference Electrode
Not Used
Green
Solution Earth (Ground Rod)
Yellow
Screen/Shield
Clear
Glass/Metal Electrode
6 INSTALLATION
Function
Not Used
for
Diff. Input
IM/AX4PH
Issue 9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax436Ax460Ax466Ax468

Table of Contents