6.1 Maintenance and inspection items
 AeroDR Battery Charger
• Clean dust on the insert table of the AeroDR Battery
Charger with a soft lint-free cloth dampened with an-
hydrous ethanol or water and wrung well.
AeroDR Detector insert table (front)
AeroDR Detector insert table (side)
 AeroDR Battery Charger2
• Clean dust on the insert table of the AeroDR Battery
Charger2 with a soft lint-free cloth dampened with
anhydrous ethanol or water and wrung well.
AeroDR Detector insert table
6 .1 .3 Disinfection of the
AeroDR Detector
If bodily fluid or blood from a patient has contaminated the
surface of the AeroDR Detector, disinfect with a soft lint-
free cloth dampened with a small quantity of the following
disinfectant and wrung well.
• Ethanol for disinfection
• Isopropanol for disinfection
• Commercial chlorine bleach, or 0.5% hypochlorite
(10-fold dilution of household bleach)
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• Bleach and hypochlorite are corrosive, so wash the
bleach off well to avoid corrosion.
• Be careful not to apply any chemical for disinfection
onto the wired connection connectors and the LEDs.
• If you directly apply or spray chemical for disinfec-
tion on the AeroDR Detector, the solution will enter
the instrument through exterior gaps, causing a
failure.
•• • • •• • • •• • • •••••••••••••••••••••••••••••••••••••••••
6 .1 .4 Consumables
IMPORTANT
•••••••••••••••••••••••••••••••••• • ••
• Refer to each device's manual for information
about periodic replacement parts and consumables
for the image processing controller, etc.
• In particular, continued use of the battery may re-
sult in degradation and wear, and it may no longer
exhibit proper functioning capabilities. For extend-
ed, safe use, it is necessary to replace parts which
have become worn or degraded.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
6 .1 .5 AeroDR Grip sheet affix-
ing and replacement
When affixing the AeroDR Grip sheet to the AeroDR De-
tector, follow the procedure below.
Application locations
AeroDR Grip sheet
•••••••••••••••••••••••••••••••••• • ••
IMPORTANT
• When using a tool to affix the AeroDR Grip sheet or
perform replacement, be careful not to get injured
by a tip or edge of the tool.
• For AeroDR Grip sheet affixing or replacement, it
is recommended to ask Konica Minolta technical
representative.
• The AeroDR 1012HQ does not use the AeroDR
Grip sheet.
••••••••••••••••••••••••••••••••••••••••••••••••• • •• •
92