5 .2 .6 AeroDR Battery Charger

Status
Status LED (green) does not
light.
Error LED (orange) is on.
When the AeroDR Detector
*1
is set, the error LED (orange)
lights.
Although the AeroDR Detector
*1
inserted, charging does not start,
and the status LED (green) is out.
Although the AeroDR Detector
is
*1
inserted, it cannot be registered, and
the status LED (green) is flashing.
*1 The AeroDR 1012HQ cannot be inserted into the AeroDR Battery Charger.
5 .2 .7 AeroDR Battery Charger2
Status
Status LED (green) does not
light.
Error LED (orange) is on.
When the AeroDR Detector is
set, the error LED (orange) lights.
Although the AeroDR Detector is
inserted, charging does not start,
and the status LED (green) is out.
Although the AeroDR Detector is
inserted, it cannot be registered,
and the status LED (green) is
flashing.
5 .2 .8 AeroDR Access Point
If a problem occurs in the AeroDR Access Point, contact Konica Minolta technical representatives.
Error description
AC power not supplied.
-
AeroDR Detector was inserted the wrong way.
The AeroDR Detector is not set correctly.
is
AC power not supplied.
The Ethernet cable is disconnected.
Error description
AC power not supplied.
The power switch is turned off.
-
AeroDR Detector was inserted the wrong way.
The AeroDR Detector is not set correctly.
AC power not supplied.
The power switch is turned off.
The Ethernet cable is disconnected.
87
5.2 Various problems and countermeasures
Corrective actions
Make sure that the power cable is connected
correctly.
Contact Konica Minolta technical
representatives.
Check the insertion orientation of the AeroDR
Detector.
Make sure that the AeroDR Detector is set
correctly.
Make sure that the power cable is connected
correctly.
Make sure that the Ethernet cable is
connected correctly.
Corrective actions
Make sure that the power cable is connected
correctly.
Turn on the power switch.
Contact Konica Minolta technical representa-
tives.
Check the insertion orientation of the AeroDR
Detector.
Make sure that the AeroDR Detector is set
correctly.
Make sure that the power cable is connected
correctly.
Turn on the power switch.
Make sure that the Ethernet cable is connect-
ed correctly.

Hide quick links: