 AeroDR Detector
• Clean the exterior and AeroDR Grip sheet with a
soft lint-free cloth dampened with a small amount of
anhydrous ethanol and wrung well.
• To clean the gaps between the exterior of the
AeroDR Detector and the protective cover, and the
gaps between the exterior of the AeroDR Detector
and the AeroDR Grip sheet, remove dirt using a com-
mercial plastic brush.
Gap between
the exterior and
AeroDR Grip sheet
AeroDR Grip sheet
Exterior
Protective cover
Gap between
the exterior and
protective cover
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• If you directly apply or spray anhydrous ethanol
on the AeroDR Detector, the liquid will enter the
AeroDR Detector through exterior gaps, causing a
failure.
•• • •• •• •• • •• •••••••••••••••••••••••••••••••••• •••••••
 Spring connector
• If foreign material has adhered to the spring con-
nectors of the AeroDR Battery Charger2, AeroDR
I/F Cable and AeroDR UF Cable, remove it with a
commercial plastic brush.
6.1 Maintenance and inspection items
AeroDR Battery Charger2
AeroDR I/F Cable
Spring connector
AeroDR UF Cable
Spring connector
 Wired connection connector
• If foreign material has adhered to the wired connec-
tion connectors of the AeroDR Detector and AeroDR
UF Cable, remove it with a commercial plastic brush.
AeroDR Detector
Wired connection connector
AeroDR UF Cable
91
Spring connector
Wired connection connector