Konica Minolta AeroDR System Operation Manual page 55

Direct digitizer
Hide thumbs Also See for AeroDR System:
Table of Contents

Advertisement

•••••••••••••••••••••••••••••••••••••
IMPORTANT
• For an AeroDR 1417S without the "1417S" identifica-
tion, use a grid with a grid density of 34 lp/cm and an
angle error of 0.5° or less.
• When the "Aero Sync mode" is used, use a grid with a
grid ratio of 8 or less.
• A capped grid is recommended when laying the grid
over the AeroDR Detector on the table top or for expo-
sure with the X-ray device.
• Do not use a grid of which the antenna is covered with
metal, as the quality of wireless communication may
be significantly reduced.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
 High-dose exposure
• When high-dose exposures are performed con-
tinuously, afterimages of the last exposure will
occasionally be visible. The affect of afterimages
in this device can be resolved by leaving longer
intervals between exposures, so make the inter-
vals between exposures longer than usual when
performing high-dose exposure.
• During high-dose exposure, continual use of a
lead or other marker at the same position can
cause burning in of the afterimage. Therefore,
avoid using it in the same position. Also, if image
burn-in is visible, run calibration.
AeroDR P-12
1417S
(
)
Identification
55
3.2 Operation of AeroDR Detector
 Exposure time for basic connection/S-SRM
connection
• Select the maximum exposure time from 0.7, 1.7,
and 3.2.
• AeroDR Detectors whose maximum exposure
time can be set to 4.0, 6.7, 10.3 are as follows.
AeroDR
Detector
First four digits
AeroDR 1417HQ A5DP
AeroDR 1417S
A50D
AeroDR 1717HQ A6C3
AeroDR 1012HQ A5TE
• The exposure time for actual exposures should
not exceed the maximum exposure time you have
set. Otherwise, exposure may fail or correct im-
ages may not be able to be obtained.
IMPORTANT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• You can select a maximum exposure time of 4.0, 6.7,
and 10.3 for the AeroDR 2 1417HQ and AeroDR 2
1417S.
• If you increase the maximum exposure time, the
AeroDR Detector may take longer to exit sleep mode.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• • • • •
HINT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• Use the image processing controller to set the maxi-
mum exposure time. In some cases, the maximum
exposure time is fixed according to the X-ray device.
For details, contact Konica Minolta technical represen-
tatives.
• When using ImagePilot as the image processing con-
troller, it is not possible to change the maximum expo-
sure time.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
 Exposure time for Aero Sync mode
• When using Aero Sync mode, set the maximum
exposure time to 0.8 seconds or less.
• When using Aero Sync mode, set the minimum
exposure time as follows:
AeroDR Detector
AeroDR 1417HQ/
AeroDR 1717HQ/
AeroDR 1012HQ
AeroDR 2 1417HQ/
AeroDR 2 1417S
Serial Number
Number
50198 or more
51001 or more
51001 or more
51001 or more
Minimum
exposure time
2.1 milliseconds
1 millisecond

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aerodr system 2

Table of Contents