1 .3 .3 Precautions regarding
electromagnetic waves
 EMC Statement
This device has been tested and found to comply with
the limits for medical devices in IEC 60601-1-2: 2007.
These limits are designed to provide reasonable pro-
tection against harmful interference in a typical medi-
cal installation. The device generates, uses and can
radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause
harmful interference to other devices in its vicinity.
However, there is no guarantee that interference will
not occur in a particular installation.
Whether this device does cause harmful interference
to other devices can be determined by turning this de-
vice off and on. If it causes harmful interference, the
user is encouraged to try to correct the interference by
1 or more of the following measures:
• Reorient or relocate the receiving device.
• Increase the separation between the devices.
• Connect this device into a wall outlet on a circuit differ-
ent from that to which the other devices are connected.
• Contact Konica Minolta technical representatives.
 Supplementary information regarding IEC 60601-
1-2: 2007
(1) Take precautions against this device especially re-
garding EMC. Install and put into service according
to the electromagnetic compatibility (EMC) informa-
tion provided in the manual (Table 1 - Table 4).
(2) Do not use mobile phones or pocket pagers in the vicin-
ity of this device. Use of mobile phones or pocket pag-
ers near this device can cause errors in operation due
to electromagnetic wave interference, so such devices
should be turned off in the vicinity of this device.
(3) Cable list
• Power cable (3.048m/3-Wire/No Shielding)
• Ethernet cable (max 100m/100 BASE-TX)
• Ethernet cable (PoE) (30m/No Shielding)
• Various AeroDR I/F Cables
• Various AeroDR UF Cables
• AeroDR Collimator Cable Set
• Various AeroDR XG Cable Sets
(4) The use of accessories, transducers and cables
other than those sold by Konica Minolta, Inc. as
internal components, may result in increased emis-
sions or decreased electromagnetic immunity of
this device.
(5) Do not use this device adjacent to or stacked with
other devices. If adjacent or stacked use is neces-
sary, confirm normal operation in the configuration
in which this device will be used.
(6) Specifications regarding RF transmitters frequency:
• Frequency
– AeroDR SYSTEM:
5150 to 5350MHz, 5470 to 5850MHz
– AeroDR SYSTEM 2:
2412 to 2472MHz
5180 to 5320MHz, 5500 to 5825MHz
• Modulation
– AeroDR SYSTEM:
OFDM
– AeroDR SYSTEM 2:
2412 to 2472MHz: DSSS/CCK/OFDM
5180 to 5320MHz, 5500 to 5825MHz:
OFDM
• Maximum effective radiation power
– AeroDR SYSTEM:
+15dBm
– AeroDR SYSTEM 2:
+10dBm
• This device may be interfered with by other devic-
es that conform to CISPR emission requirements.