– Emit the X-rays onto the entire AeroDR Detec-
tor.
– Emit the X-rays by using a tube current of 140
mA or higher or a tube voltage of 90 kV or
higher.
•••••••••••••••••••••••••••••••••••••
IMPORTANT
• If the conditions are not met, X-rays may not be de-
tected or streaks may occur in the image.
•• • •• •• •• • •••••••••• ••••••••••••••••••••••••••••••••
 S-SRM connection
• When S-SRM connection is adopted, make sure
to perform exposure using the hand switch con-
nected to the AeroDR Interface Unit2, AeroDR
Generator Interface Unit and AeroDR Generator
Interface Unit2.
• If exposure is performed using the hand switch on
the operation panel of the X-ray device, exposure
synchronized to the AeroDR SYSTEM is not pos-
sible.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• If you restart the image processing controller, also
restart the AeroDR Interface Unit or AeroDR Interface
Unit2.
•• • •• •• • •• •• •••••••••••••••••••••••••••••••••••••••••
 AED (Automated External Defibrillator)
• When an AED (Automated External Defibrillator)
is used, move the AeroDR Detector away from
the patient. High voltage and high current may
result in a breakdown of the AeroDR Detector.
 The operating temperature environment of
AeroDR Detectors
• If an AeroDR Detector is left close to or in contact
with a heat generator such as electric carpet, its
safety device may be activated and the AeroDR
Detector may fail to work properly.
3.2 Operation of AeroDR Detector
 Waterproof performance of the AeroDR 2
1417HQ/AeroDR 2 1417S
• If there is a risk of the ingress of patient body
fluids, chemicals, water, etc., make sure to cover
the AeroDR 2 1417HQ and AeroDR 2 1417S with
a vinyl sheet, etc. The ingress of patient body
fluids, chemicals, water, etc. into the protective
cover cannot be prevented. In particular, if patient
body fluids enter into the protective cover, for hy-
gienic purposes, make sure to replace or clean
the parts.
• Subjecting the AeroDR 2 1417HQ and AeroDR
2 1417S to strong impacts or excessive loads
by dropping or hitting them against other objects
may diminish their waterproof capabilities.
• The AeroDR 2 1417HQ and AeroDR 2 1417S are
designed to have IPX6 waterproof capabilities,
but they are not guaranteed to be completely wa-
terproof.
IMPORTANT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• The AeroDR 1417HQ, AeroDR 1417S, AeroDR
1717HQ, and AeroDR 1012HQ are not waterproof.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
HINT
• ••• •• •• •• ••• •• •• •• •• ••• •• •• •• ••• • • • •
• An IPX6 (level of jet stream protection) rating indicates
that the AeroDR 2 1417HQ and AeroDR 2 1417S
are designed to operate even after being exposed to
normal temperature water projected from all angles
through a 12.5mm nozzle for 3 minutes or more (at a
flow rate of about 100L/min) from a distance of 3 me-
ters.
• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• ••• •• •• •• •• ••• •• •• • • • • •
57
Protective cover