Status
The AeroDR I/F Cable cannot
be connected to the AeroDR
Detector.
Only wired connection with the
AeroDR Detector cannot be used.
Only wireless connection with
the AeroDR Detector cannot be
used.
AeroDR Detector wired connec-
tion connector is hearted up.
Charging sometimes takes lon-
ger.
Battery LED (blue) flashes fast.
*2
No image is acquired.
Reading has automatically
*2
started.
Error description
Part of the wired connection connector of the
AeroDR Detector is deformed.
The spring connector of the AeroDR I/F Cable
is deformed.
Foreign material is in the wired connection
connector of the AeroDR Detector.
Foreign material is in the spring connector on
the AeroDR I/F Cable.
The wired cable is not connected properly.
There is an error in the access point.
The AeroDR Detector and access point are
being used under poor conditions.
• Wireless does not connect
• Wireless communication terminates
• Cycle time is extended
AeroDR Detector wired connection connector
is heated up immediately after charging with
AeroDR Battery Charger
*1
or AeroDR Battery
Charger2.
Charging may take some time if the battery is
completely discharged.
Usage time with the battery has gotten shorter.
The number of images that can be exposed has decreased.
The charging time has gotten shorter.
No image appears after X-ray exposure.
The machine starts reading the image without
X-ray exposure.
83
5.2 Various problems and countermeasures
Corrective actions
Contact Konica Minolta technical
representatives.
Refer to "6.1.2 Cleaning" and remove the
foreign material.
Check that the wired cable is properly con-
nected to the AeroDR Detector.
Check that the Ethernet cable is properly con-
nected to the access point.
Check the installation location of AeroDR
Detector and access point.
When performance is noticeably lower than at
the initial installation of this device, it is pos-
sible that the installation environment or usage
environment have changed.
This is caused by charging and is not a mal-
function.
It takes time for internal components to start.
Because it is not abnormal, wait a bit.
It could be that the charging function of the
battery has deteriorated. It can be replaced
with a new battery for a fee if necessary.
(1) Check the following items.
• Did you perform exposure in "Ready" status?
• Are the exposure conditions correct?
• Was X-ray output lowered?
• Was X-ray voltage set to a lower value?
(2) Perform the following procedures.
• For AeroDR 1417HQ, AeroDR 1417S,
AeroDR 1717HQ and AeroDR 1012HQ
– Set the sensitivity setting of the image
processing controller to "High" before
exposure. If the setting is already set to
"High", increase the voltage of the X-ray
and perform exposure.
• For AeroDR 2 1417HQ and AeroDR 2 1417S
– Increase the voltage of the X-ray before
exposure.
(1) Check the following items.
• For AeroDR 1417HQ, AeroDR 1417S,
AeroDR 1717HQ and AeroDR 1012HQ
– Was the AeroDR Detector subjected to
strong shocks or vibrations while "Ready"
was displayed?
– Is the sensitivity setting of the image
processing controller correct? (Is the
sensitivity setting of the image processing
controller set to [High]?)
• For AeroDR 2 1417HQ and AeroDR 2 1417S
– Was the AeroDR Detector subjected to
strong shocks or vibrations while "Ready"
was displayed?
(2) Use caution to prevent strong shocks or
vibrations and restart exposure.