1 .3 .5 Precautions for installing,
moving, and storing
WARNING
• Take note of the following when moving this device
not including the AeroDR Detector:
– Do not subject to shocks or vibration .
– Do not start moving until the power is turned off,
and operation has stopped completely.
– Do not move with the power cable or any other
cable connected.
• When installing the AeroDR Battery Charger or AeroDR
Battery Charger2 on a table top, observe the following:
– Install on a base that can withstand maximum
patient weight because load is inserted when
setting the AeroDR Detector.
– The contact area for the AeroDR Battery Charger
or AeroDR Battery Charger2 must be larger than
the area of the AeroDR Battery Charger or AeroDR
Battery Charger2 so that it will not fall down when
the AeroDR Detector is set or when people touch it
by accident.
– Install on a horizontal and stable surface.
– Do not use a base that has a surface made of
slippery material such as teflon.
– Connect power cable and wires so that no one will trip.
– Install in a place easy to set the AeroDR Detector.
– Install so that the vent on the back side of the
AeroDR Battery Charger is not blocked.
CAUTION
• Because connections of the X-ray device can only be
made by Konica Minolta or its designated contractors,
contact Konica Minolta or its designated contractors.
• For the X-ray devices enabled to connect, contact
Konica Minolta technical representatives.
• Contact Konica Minolta or dealers specified by
Konica Minolta to install or move this device.
• Since it is required to meet the specification pro-
vided by the X-ray device manufacturer to connect
with an X-ray device, contact Konica Minolta or
dealers specified by Konica Minolta.
• Take note of the following when installing or storing
this device.
– Install or store this device within the specified
storage and usage conditions and environment.
– Do not install or store in a location where it may
be adversely affected by atmospheric pressure,
temperature, humidity, ventilation, sunlight, dust,
salt-air, or air containing sulfur.
– Do not install or store in a location where it is not
stable, ventilation is insufficient, the difference
in light-dark is great, electromagnetic waves are
generated, or where subject to vibration or shock.
– Do not install or store in a location where chemi-
cal agents are used or stored.
– Do not install this device facing up or upside down.
– Do not install the AeroDR Interface Unit, AeroDR
Interface Unit2 and AeroDR Generator Interface
Unit stacked on each other.
• Connect the AeroDR Interface Unit2, AeroDR Gen-
erator Interface Unit and AeroDR Generator Inter-
face Unit2 to an X-ray device that conforms to IEC
60601 or to an equivalent standard.
• Please see "8.4 Compatible X-ray system specifi-
cation".
1 .3 .6 Precautions regarding
maintenance
• Perform the maintenance and inspection periodi-
cally. In addition to the user periodical maintenance
that needs to be performed, periodical maintenance
by a service engineer is also required.
• If there are stains such as body fluids, clean and
disinfect.
• Based on the warranty, parts that are no longer un-
der warranty (1 year) can be replaced for a fee.
• Turn off the power and disconnect the power plug
from the wall outlet before cleaning or maintaining
this device.
• Securely connect the power cable, AeroDR I/F
Cable, and AeroDR UF Cable after cleaning and
maintenance.
• Clean the dirt from between the protective cover
and the exterior, and the terminal using a commer-
cial plastic brush. Do not clean with sharp or hard
objects.
• Take care regarding the following when disinfecting
the AeroDR Detector.
– Use ethanol for disinfection, isopropanol for
disinfection, or commercial chlorine bleach, or
0.5% hypochlorite (10-fold dilution of household
bleach) when disinfecting. However, bleach and
hypochlorite are corrosive, so wash the bleach
off well to avoid corrosion.
25