5.2 Various problems and countermeasures

5 .2 .3 AeroDR Interface Unit2

Status
Power LED (green) does not
light.
Feeding1/2 LED (blue) does not
light.
Communications are down
between devices, and power
LED (green) is off.
The AeroDR Detector is being
used with wired connection,
but it is not recognized on the
image processing controller, and
feeding1/2 LEDs (blue) are off.
Battery is not charging while
the AeroDR Detector is used
with wired connection (the
battery level on the battery level
display of the image processing
controller does not increase),
and feeding1/2 LEDs (blue) are
off.
5 .2 .4 AeroDR Generator Interface Unit
Status
LED (green) does not light.
LED (green) does not change
from flashing to a solid light.
5 .2 .5 AeroDR Generator Interface Unit2
Status
LED (green) does not light.
LED (green) does not change
from flashing to a solid light.
Error description
AC power not supplied.
Feeding1/2 LED (blue) does not light even
when connecting the AeroDR Detector.
The power switch is not on. Or, the power
cable is not connected.
The AeroDR I/F Cable is not connected.
Error description
AC power not supplied.
There is a communications error.
Error description
AC power not supplied.
There is a communications error.
Make sure the power cable is connected cor-
rectly.
Make sure that the AeroDR I/F Cable is con-
nected correctly.
Make sure the power switch of the AeroDR
Interface Unit2 is on. Or, make sure the power
cable is connected correctly.
Make sure that the AeroDR I/F Cable is
connected to the AeroDR Detector properly.
Make sure that the Ethernet cable is
connected correctly. Or, make sure the power
switch of the AeroDR Interface Unit is on.
Contact Konica Minolta technical
representatives.
Make sure that the Ethernet cable is
connected correctly. Make sure the power
switch of the AeroDR Interface Unit or
AeroDR Interface Unit2 is on.
If the AeroDR Generator Interface Unit2 has
a power switch, make sure the power switch
is on.
Contact Konica Minolta technical
representatives.
86
Corrective actions
Corrective actions
Corrective actions