5.2 Various problems and countermeasures
5 .2 .9 Image processing controller/Images
Status
Transverse (noise) has gotten
into all images acquired from the
AeroDR Detector.
Operation is normal, but prob-
lems can be seen with exposure

images.

Exposure cannot be confirmed in
image processing controller.
Error description
The wired cable is not connected properly.
It has been occurring frequently since a cer-
tain time.
There is trouble in only 1 image.
The image processing controller does not
become Ready.
The icons of devices used on the image pro-
cessing controller are not displayed.
Connect the wired cable spring connector
unit horizontally to the wired connector of the
AeroDR Detector.
Perform calibration.
Check the exposure method and image pro-
cessing.
Check the image processing controller start
up.
Confirm that the icon of the device to be used
is displayed on the system monitor screen.
When the icon of the device is not displayed,
check that the device to be used is started. Or,
check that the Ethernet cable is properly con-
nected.