Advertisement

5 .2
Various problems and countermeasures
If the following problems occur with any of these devices, consult the respective references for countermeasures.
IMPORTANT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• After performing countermeasures, if the problem does not go away, contact Konica Minolta technical representatives.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
HINT
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
• When an error message has been displayed in the image processing controller, check the error description and countermea-
sures listed in the "Operation Manual" of the image processing controller.
• When using a general-purpose hub or access point, refer to its operation manual.
• • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • •
5 .2 .1 AeroDR Detector
Status
The AeroDR Detector does not
start up.
The AeroDR Detector does not
shut down.
The status LED (green) is lit,
and the busy/error LED (orange)
flashes rapidly.
("Ready" is not displayed on the
image processing controller)
When the AeroDR Detector is
placed on a smooth surface, the
AeroDR Detector is not stably at-
tached to the smooth surface.
The AeroDR Detector will not go
into the wall stand or table.
The AeroDR Detector cannot be
inserted in the AeroDR Battery
Charger
or AeroDR Battery
*1
Charger2.
Error description
Power does not go on even when the power
switch is pressed for 2 seconds or longer.
Power does not go off even when the power
switch is pressed for 5 seconds or longer.
System error is occurring.
AeroDR Detector is warped.
The protective cover is deformed.
AeroDR Detector is warped.
The protective cover is deformed.
The protective cover is deformed.
Foreign material is in the wired connection
connector of the AeroDR Detector.
Foreign material is in the bottom of the
AeroDR Battery Charger or AeroDR Battery
Charger2.
The battery might be out. Charge it as follows.
Then, start it.
• Inserting it into the AeroDR Battery Charger
for more than 2 minutes
• Inserting it into the AeroDR Battery Charger2
for more than 6 minutes
• Connecting a wired cable for more than 6
minutes
It is not possible to shut down while exposing.
Shut down after exposure is ended.
If a busy/error LED (orange) continues to flash
after 10 minutes has elapsed, shut down the
AeroDR Detector.
Or, when "Ready" is not displayed on the im-
age processing controller, restart it.
If the AeroDR Detector is still warped even
after the detector is placed on a smooth
surface, contact Konica Minolta technical
representatives.
Contact Konica Minolta technical
representatives.
If the AeroDR Detector is still warped even
after the detector is placed on a smooth
surface, contact Konica Minolta technical
representatives.
Contact Konica Minolta technical
representatives.
Refer to "6.1.2 Cleaning" and remove the
foreign material.
Contact Konica Minolta technical
representatives.
82
Corrective actions