3 .2
Operation of AeroDR Detector
3 .2 .1 Exposure
Exposure with this device is performed with the follow-
ing procedure.
 Exposure under the basic connection ex-
ample
1
Perform examination registration with the
image processing controller .
2
Check that this device is ready to expose
images, and then prepare for the expo-
sure .
3
Push the exposure switch of the X-ray de-
vice to perform the exposure .
• When the exposure is completed, images are
stored in the AeroDR Detector and will then be
converted to digital data and sent to the image
processing controller sequentially.
4
Check that the exposed image is dis-
played on the image processing control-
ler .
 Exposure under the S-SRM connection ex-
ample
1
Perform examination registration with the
image processing controller .
2
Check that this device is ready to expose
images, and then prepare for the expo-
sure .
3
Push the S-SRM's hand switch to the first
stage .
• The exposure preparation signal is sent to the
X-ray device.
First stage switch
4
Push the S-SRM's hand switch up to the
second level to perform exposure .
• Exposure is performed from the X-ray device to
produce X-ray images.
• When the exposure is completed, images are
stored in the AeroDR Detector and will then be
converted to digital data and sent to the image
processing controller sequentially.
Second stage switch
5
Check that the exposed image is dis-
played on the image processing control-
ler .
IMPORTANT
•••••••••••••••••••••••••••••••••• • ••
• The AeroDR Detector is precision equipment, and
therefore impact or vibration during radiography or im-
age transfer may affect the image quality. Be careful
when handling the AeroDR Detector during and just
after radiography.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
•••••••••••••••••••••••••••••••••• • ••
HINT
• If the AeroDR Detector remains unused for a long time
(time can be set) it transitions to the sleep mode.
• When the image processing controller is ready to ex-
pose, it recovers from the sleep mode.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
Reference
•••••••••••••••••••••••••••••••••• • ••
• Regarding the operation of the image processing con-
troller, refer to the "Operation Manual" of the image
processing controller.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••