Maintenance And Inspection Items - Konica Minolta AeroDR System Operation Manual

Direct digitizer
Hide thumbs Also See for AeroDR System:
Table of Contents

Advertisement

6 .1

Maintenance and inspection items

This chapter describes the inspections and cleaning
required in order to maintain the use of this device in
an optimum condition.
 Quality Tool
HINT
•••••••••••••••••••••••••••••••••••••
• "Simple Check QC for CS-7" is available for check-
ing of imaging quality. Please contact our customer
service for the detail.
• For details, contact Konica Minolta technical repre-
sentatives.
• • •• • • •• • • •• • ••••••••••••••••••••••••••••••••••••••••
6 .1 .1 Maintenance schedule
The maintenance and inspection items that the user
should perform are as follows.
Maintenance task
Checking and cleaning the surface of the
AeroDR Detector
Checking for external damage to the
AeroDR Detector
Cleaning the spring connectors of the
AeroDR Battery Charger2, AeroDR I/F
Cable and AeroDR UF Cable
Cleaning the wired connection connec-
tors of the AeroDR Detector and AeroDR
UF Cable
Cleaning the AeroDR Battery Charger
and AeroDR Battery Charger2
Full charge of the AeroDR Detector
Calibration
•••••••••••••••••••••••••••••••••••••
IMPORTANT
• To ensure optimum use of this device, be sure to
perform periodic maintenance.
• The above task intervals are estimates and vary
according to usage.
• • • •• • • •• • •• •••••••••••••••••••••••••••••••••••••••••
The cleaning methods of the respective devices are as
follows.
Maintenance
interval
Weekly
Weekly
Weekly
Weekly
Weekly
Monthly
Every 3 months
or when a mes-
sage is displayed
90
HINT
••••••••••••••••••••••••••••••••••• • •
• Perform calibration when a message that prompts
for calibration of the image processing controller is
displayed.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
Reference
••••••••••••••••••••••••••••••••••• • •
• For the calibration, refer to the "Operation Manual"
or the "User Tool Operation Manual" of the image
processing controller.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••
6 .1 .2 Cleaning
IMPORTANT
•••••••••••••••••••••••••••••••••• • ••
• Be careful not to apply any cleaning chemical or
liquid onto the spring connectors, the wired con-
nection connectors, and the LEDs.
• Do not clean with sharp or hard metal objects. If you
cannot remove stains, contact Konica Minolta techni-
cal representatives.
• Wear and deformation of the protective cover will
occur with the AeroDR Detector due to the way it
is handled. The protective cover can be replaced
for a fee when the damage becomes extensive, so
contact Konica Minolta technical representatives.
•••••••••••••••••••••••••••••••••••••••••••••••••• • ••

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aerodr system 2