Flow Never Becomes Ready
Agilent 7890 Series Troubleshooting
If the gas flow never becomes ready, check for the following:
• Check the supply gas for
• Check the restrictors installed in the Aux EPC module. See
Restrictors
in the Advanced User Guide.
• Check the configured gas type. The configured gas type must
match the actual gas plumbed to the GC.
• Check for leaks in the gas delivery plumbing and the GC. (See
"Checking for
Leaks".)
GC Not Ready Symptoms
sufficient delivery
pressure.
4
91