Advertisement

FID Ignitor Does Not Glow During Ignition Sequence

WA RNING
Agilent 7890 Series Troubleshooting
Keep body parts at a safe distance from the FID chimney while
performing this task. If using hydrogen, the FID flame will not be
visible.
1
Remove the detector top cover.
Turn the FID flame On.
2
3
Observe the ignitor plug though the FID chimney. The small
hole should glow during ignition sequence.
If the test fails, check for the following possible causes:
• The ignitor may be bad; replace the ignitor. (See
Maintenance on the FID Collector
• Detector temperature is set to < 150 °C. Agilent recommends
operating the FID at > 300 °C.
• The ignitor is not making a good connection to the ground:
The ignitor must be tightly screwed into the FID castle
assembly.
The three T-20 Torx screws that hold the collector
assembly in place must be tight.
ALS and Detector Symptoms
To Perform
Assembly.)