To Use The Stopwatch - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

To Use the Stopwatch

Agilent Technologies
In the stopwatch mode, both the time (to 0.1 second) and
reciprocal time (to 0.01 min—1) are displayed. Use the
stopwatch when measuring flows with a bubble flowmeter.
1
Press [Time] and scroll to the time = line.
2
Press [Enter] to start the stopwatch.
3
Press [Enter] again to stop.
4
Press [Clear] to set to zero.
You can access other functions while the stopwatch is running.
Press [Time] again to view the stopwatch display.