Peak Fronting - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

3
Chromatographic Symptoms

Peak Fronting

56
The figure below shows examples of the three types of peaks:
symmetric, fronting, and overloaded.
Symmetrical
If peak fronting or overloading occurs, try the following:
• Verify that the injection volume is appropriate. Column
overload is the most common cause for fronting peaks.
Decrease injection volume, dilute the sample, and/or
increase the split ratio.
Change the column type or film thickness.
• Ensure that the column is
the column in the inlet.
• Verify that the appropriate injection technique is being used.
This is usually related to erratic plunger depression or
having sample in the syringe needle.
• If using capillary splitless injection, consider the compound
solubility in the injection solvent.
Change the solvent.
Use a retention gap.
• Check purity of sample solvent. For solvents with large
differences in polarity or boiling points, a mixed sample
solvent can cause peak fronting. Change the sample solvent.
Fronting
Overloaded
installed
properly. If not, reinstall
Agilent 7890 Series Troubleshooting