GC Not Ready Symptoms

GC Never Becomes Ready

90
Normally the GC becomes ready after flows and temperatures
reach setpoint. If the GC does not become ready after a long
period of time:
• Press [Status] or a component key (for example, [Front inlet])
to see which setpoints or conditions are not ready.
• Check for a sampler problem.
• Check for a data system problem.
• If performing manual injections in splitless or gas-saver
mode, you may need to press [Prep Run] to prepare the inlet
for the injection. Do this, for example:
To toggle the inlet purge valve before a splitless injection
To prepare for a pulsed injection
To turn off gas saver.
For more information on [Prep Run], see the
Advanced User Guide.