Fid Does Not Ignite - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

FID Does Not Ignite

Agilent 7890 Series Troubleshooting
• Verify that the Lit Offset is <
• Ensure that the FID temperature is high enough for ignition
(>150 °C). Agilent recommends >300 °C.
• Check that the FID ignitor glows during ignition sequence.
(See
To Verify FID Ignitor Function During Ignition
Sequence.)
• Check that the air and hydrogen pressures meet Agilent's
recommendations (hydrogen > 35 psi [210 kPa] and air > 55
psi [380 kPa]). See the
Preparation
Guide.
• Try increasing the supply pressures to the FID flow module.
This makes the flame easier to light without changing the
setpoints.
• Increase hydrogen and air flow rates until ignition occurs,
then reduce them toward the method values. Experiment for
the best values.
• Check for a plugged or partially plugged jet. (See
for a Plugged FID
Jet.)
• Measure the FID flow rates. Actual flow rates should be
within +/-10% of the setpoint. (See
The hydrogen:air ratio greatly impacts ignition. Nonoptimal
flow settings can prevent flame ignition. (See
Detector
Flow.)
• There could be a large leak in the system if the flame still will
not light. Large leaks result in measured flow rates being
different from actual flow rates, causing nonideal ignition
conditions. Thoroughly leak check the whole system,
especially the column fitting at the FID. (See
Leaks).
• Check the column flow rate. (See
Flow.) Hydrogen flow should be greater than the sum of the
column flow and makeup flow.
• If the analysis permits, substitute nitrogen for helium as
makeup.
GC Not Ready Symptoms
2.0 pA.
Agilent GC, GC/MS, and ALS Site
FID starting
To Measure a Column
4
To Check
conditions.)
To Measure a
Checking for
101

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7890a

Table of Contents

Save PDF