Retention Times Not Repeatable - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

3
Chromatographic Symptoms

Retention Times Not Repeatable

42
• Replace the septum.
• Check for leaks in the inlet, liner (as applicable), and column
connection. (See
"Checking for
To Perform an Inlet Leak Check
To Check for Leaks in a Split/Splitless Inlet
To Perform a PP Pressure Decay Leak Test
To Check for Leaks in a Multimode Inlet
To Perform a COC Pressure Decay Leak Test
To Perform a PTV Pressure Decay Test
To Perform a VI Pressure Decay Test
• Check for sufficient carrier gas supply pressure. The
pressure delivered to the GC must be at least 40 kPa (10 psi)
greater than the maximum inlet pressure required at final
oven temperature.
• Run replicates of known standards to verify the problem.
• Verify that you are using the correct liner type for the sample
being injected. (See
• Consider if this is the first run. (Has the GC stabilized?)
• If using an FID or NPD and retention times increase (drift),
check the jet for contamination or replace the jet.
To Replace an FID Jet
To Check for a Plugged FID Jet
To Maintain the NPD Collector, Ceramic Insulators, and
Jet
To Check for a Plugged NPD Jet
Leaks".)
Selecting the correct inlet
Agilent 7890 Series Troubleshooting
liner.)