Vial Mishandled by ALS (7693A)

Agilent 7890 Series Troubleshooting
Refer to your
sampler operating documentation
information.
When you find a mishandled sample vial, do the following:
• Check for folds or wrinkles in the crimp cap, especially near
the neck of the sample vial.
• Use Agilent-recommended sample vials.
The figure below shows the critical dimensions for sample
vials and microvial inserts to be used with the 7693A ALS
system. These dimensions do not make up a complete set of
specifications.
Body Diameter (BD) = 11.7 ± 0.2
Cap Diameter (CD) = BD × 1.03 maximum
All dimensions in millimeters
CD
33.5 ± 1.5
Vial plus cap height
BD
Maximum height of a capped vial
• Verify the transfer turret is installed if using a G4514A tray.
• Check sample labels (if applicable).
Check that they are the correct size.
Verify that the labels do not interfere with the gripper.
Vials are available with a write-on spot for easy marking. If
you choose to make and apply your own labels, Agilent
Technologies recommends the positioning and maximum
label thickness shown in the figure below.
ALS and Detector Symptoms
for additional
6.25 ± 0.25
3.7 min.
8.2 ± 0.2
32 ± 0.5
11.7 ± 0.2
Crimp cap sample vial
2
25 minimum
body height
19