To Correct Leaks In The Cool On-Column Inlet - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

7
Checking for Leaks

To Correct Leaks in the Cool On-Column Inlet

160
If the inlet fails a pressure decay test, check the following:
• Check the caps/plugs used in the test—make sure each is
correctly installed and tight.
• If you performed the leak test after performing maintenance,
check for proper installation of the part(s) handled during
the maintenance.
• Check the tightness of the septum nut or cooling tower
assembly. See
To change a septum nut or cooling tower and
septum on a COC
inlet.
• Check the septum. Replace if old or damaged.
• Make sure the inlet temperature remained constant during
the test.
If these items do not resolve the problem, contact Agilent for
service.
Agilent 7890 Series Troubleshooting

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7890a

Table of Contents