To Check a Split Vent Line for Contamination

WA RNING
192
This procedure applies to the multimode and split/splitless
inlets.
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
1
Gather the following:
7/16-inch wrench
Wrench, hex for changing septum
Heat-resistant gloves if the inlet is hot
Cleaning brushes
2
Load the
inlet maintenance method
become ready.
3
Loosen the 1/8-inch Swagelok nut that seals the copper split
vent line to the injection port.
4
Inspect the 1/8-inch copper split line for condensed sample.
Flush or replace the copper split vent line.
5
Remove the septum head.
6
Check the split tube fitting in the shell weldment for
blockage. Clean the split tube with a brush and suitable
solvent.
7
Reassemble the split/splitless inlet and run the split vent
restriction test to test for restrictions in the split vent flow
path. If the inlet still fails, suspect flow module
contamination or defects.