To Check For Leaks In A Multimode Inlet; Cannot Reach Pressure Setpoint - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

7
Checking for Leaks

To Check for Leaks in a Multimode Inlet

Cannot Reach Pressure Setpoint

144
This procedures describes how to check for and fix leaks in the
Multimode inlet. Follow the procedures below based on the
inlet's symptoms.
If the Multimode EPC inlet does not reach its pressure setpoint,
the GC will be "Not Ready." Pressing [Status] indicates the front
(or back) inlet pressure is not ready. The GC will shut down in
approximately 5.5 minutes if the inlet cannot pressurize and
control.
If you have recently performed maintenance, first check for
leaks on any fittings/parts you handled.
1
Verify sufficient gas supply pressure delivered to the GC (see
the
GC, GC/MS, and ALS Site Preparation
no leaks in the gas supply (see
The inlet requires 70 kPa (10 psi) above the highest pressure
used in the method.
120 psi maximum for 0–100 psi inlet
170 psi maximum for 0–150 psi inlet
2
Check the total flow setting. Total flow must be high enough
to maintain inlet pressure throughout the run. Wide bore
columns require higher flow rates. Typically 50 mL/min is
sufficient. To increase total flow:
Increase split ratio if in split mode
Increase the purge flow for splitless mode
3
Put your thumb (or a septum) over the split vent exhaust. If
the inlet pressure begins to rise to the setpoint, call Agilent
for service. If the pressure remains low, continue with the
next step.
4
Perform the inlet leak check. See
Check". The total flow provides an indication of the
magnitude of the leak. Monitor the total flow while
checking/tightening fittings:
Septum nut
Column
Split vent trap/O-rings
Liner/O-rings
Split vent line connection to the inlet body
Guide), and verify
To Check for External
"To Perform an Inlet Leak
Agilent 7890 Series Troubleshooting
Leaks).