To Perform a PTV Pressure Decay Test

To Perform a PTV Pressure Decay Test

Agilent 7890 Series Troubleshooting
See the procedure below.
If the test fails:
• Check the plugged column fitting and septum purge cap.
• Check the septum head. Tighten if needed.
Replace
the septum if necessary.
• If using a septumless head, tighten.
• Inspect and
re-install
Replace
the inlet adapter if it is leaking.
• Tighten the split vent trap if needed, or
cartridge and O-rings.
• Tighten the split vent line connection to the inlet.
The pressure decay test checks for leaks from the inlet flow
module up to the column fitting.
After performing maintenance, first check for leaks in
externally accessible areas. See
If a leak is known to exist, check the externally accessible inlet
fittings first, especially any connection that has seen recent
maintenance, such as the septum nut, column adapter, column
connection, and so forth.
The pressure decay leak test described below requires removing
the column and capping the inlet column fitting. This test
can/cannot find the following types of leaks:
Checking for Leaks
Rebuild
the liner if it is leaking.
install
"To Check for External
7
as needed.
a new
Leaks".