To Measure FID Leakage Current

Agilent 7890 Series Troubleshooting
1
Load the analytical method.
Make sure flows are acceptable for ignition.
Heat the detector to operating temperature or to 300 °C.
2
Turn off the FID flame.
Press [Front Det] or [Back Det], then scroll to Output.
3
4
Verify that the output is stable and < 1.0 pA.
If the output is unstable or > 1.0 pA, turn off the GC and
check for proper assembly of the upper FID parts and for
contamination. If the contamination is confined to the
detector,
bakeout the
5
Turn on the flame.
Troubleshooting Tasks
FID.