5
Shutdown Symptoms
Clearing an MS Shutdown
After resolving an MS Shutdown
118
Unlike other shutdowns, you cannot clear this state by turning
off a setpoint (because setpoint changes are blocked). Instead,
clear this state by disabling GC to MS communications.
5977A, 7000C, 7010:
Press [MS/Aux Det].
1
Scroll to MS Communication.
2
Press Off/No.
3
5977B:
Press [Config][MS/Aux Det].
1
Scroll to Lvds communication.
2
Press Off/No.
3
If the GC started the shutdown, you can now troubleshoot the
problem and resolve the GC fault.
After resolving the GC or MS problem, always restore GC to MS
communications.
5977A, 7000C, 7010:
Press [MS/Aux Det].
1
Scroll to MS Communication.
2
Press On/Yes.
3
5977B:
Press [Config][MS/Aux Det].
1
Scroll to Lvds communication.
2
Press On/Yes.
3
Agilent 7890 Series Troubleshooting