To Adjust the FPD Lit Offset

Agilent 7890 Series Troubleshooting

To adjust the FPD Lit offset:

Press [Config].
1
Scroll to Front detector or Back detector (wherever the detector
2
is installed) and press [Enter].
Scroll to Lit offset. With the Lit offset line highlighted, enter the
3
new parameter for the FPD (typical value is 2.0 pA), and
press [Enter].
Troubleshooting Tasks
205