To Perform a SS Split Vent Restriction Test

Agilent 7890 Series Troubleshooting
Use the Agilent Instrument Utilities Split Vent Restriction Test
to determine if there is an abnormal restriction in the split vent
flow path. Open the software, select the GC, and run the test for
the inlet. (Alternately, see
If the test fails:
• Consider whether the setup exceeds the
pressures
obtainable by the GC. See also
• Consider the liner used. Agilent
pressure-drop liner for split analyses, part number
5183-4647.
Replace
the split vent trap filter.
• Inspect and
replace
pressure drop liner, consider installing a high-flow gold seal,
5182-9652.
• Check for contamination in the split vent line. See
a Split Vent Line for
If the problem remains, contact Agilent for service.