Peak Tailing - Agilent Technologies 7890 Series Troubleshooting Manual

Gas chromatograph
Hide thumbs Also See for 7890 Series:
Table of Contents

Advertisement

3
Chromatographic Symptoms

Peak Tailing

52
The figure below shows an example of tailing peaks. When
troubleshooting tailing peaks, consider:
• Which peaks are tailing?
• Are the tailing peaks active compounds, all compounds, or
are there trends (such as early eluters or late eluters)?
• Check the column for severe contamination.
If using a capillary column,
the front of the column.
For bonded and cross-linked phases, solvent rinse the
column.
Check for inlet contamination. Tailing will sometimes
increase with compound retention. Clean the inlet and
replace contaminated inlet parts.
Maintenance manual.)
• Consider the column stationary phase (active column). This
only affects active compounds. An active column usually
produces tailing that increases with retention time.
Cut off
1 meter from the front of the column.
Replace the column.
• Verify that the column was cut and installed properly.
Re-cut and
reinstall the column
the ferrules. Make a clean, square cut using a reliable tool.
Confirm the installation is leak free. If there is a leak at
the column fitting, you will see more tailing for early
eluting peaks. (See
• Consider the type of adapter, liner, and inlet seal being used.
One or all of these may be contaminated or active.
Use a new, deactivated liner. This only affects active
compounds.
remove
1/ 2 to 1 meter from
(See the 7890 Series
into the inlet and replace
"Checking for
Leaks".)
Agilent 7890 Series Troubleshooting

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7890a

Table of Contents